The search functionality is under construction.
The search functionality is under construction.

Author Search Result

[Author] Kazuma MOTOMURA(1hit)

1-1hit
  • A Microwave Doppler Radar System for Noncontact Measurement of Head and Finger Movements in Clinical Use

    Ikuo ARAI  Kazuma MOTOMURA  Tsutomu SUZUKI  

     
    PAPER-Radar Applications to Industrial World

      Vol:
    E76-B No:10
      Page(s):
    1318-1324

    A method to measure the displacement from the phase rotation of the Doppler signal including the displacement information of the moving body is proposed, where the displacement resolution can be improved 4 times by making the phase rotation faster. Furthermore, this test system is applied in clinical use. The test system is built up by using a two-phase microwave Doppler sensor covering a 10GHz band, where the Doppler frequency is multiplied 4 times by signal processing. Thus, the resolution is improved from a conventional 12.6mm (in case of 11.9GHz) to 3.15mm, and practical utilization has been attained. The microwave Doppler radar system described in this paper is adequate for the displacement measurement for a relatively fast moving body. As a medical sensor for clinical use, measurement examples of head movement in a vestibule examination (vestibule oculomotor reflexive inspection) and finger movement in a cerebellum function test are given. Furthermore by using two sets of this Doppler radar system, a 2-dimensional measurement of head movement is possible.