1-2hit |
Satoru KIKUCHI Kazuyuki SAITO Masaharu TAKAHASHI Koichi ITO Hiroo IKEHIRA
This paper presents the computational electromagnetic dosimetry inside an anatomically based pregnant woman models exposed to electromagnetic wave during magnetic resonance imaging. The two types of pregnant woman models corresponding to early gestation and 26 weeks gestation were used for this study. The specific absorption rate (SAR) in and around a fetus were calculated by radiated electromagnetic wave from highpass and lowpass birdcage coil. Numerical calculation results showed that high SAR region is observed at the body in the vicinity of gaps of the coil, and is related to concentrated electric field in the gaps of human body such as armpit and thigh. Moreover, it has confirmed that the SAR in the fetus is less than International Electrotechnical Commission limit of 10 W/kg, when whole-body average SARs are 2 W/kg and 4 W/kg, which are the normal operating mode and first level controlled operating mode, respectively.
Shinji TSUJI Ryuta TAKAHASHI Takeshi KATO Fumihiko UCHIDA Satoru KIKUCHI Toshinori HIRATAKA Masato SHISHIKURA Hiroaki OKANO Tsuneo SHIOTA Satoshi AOKI
Precise direct mounting of laser diode (LD) and photodiode (PD) chips on silica planar lightwave circuits (PLCs) has been investigated for application to transceiver modules. To achieve submicron optical alignment, self-aligned index marks on the PLCs and LDs were directly detected by transmission infrared light. The repeatability of the positioning was measured to be within 0.125 µm. The output power of the resultant module was 0.2 mW at 80 mA. A waveguide-type PD was also mounted in the same way, and module sensitivity of 0.25 A/W was demonstrated.