A method for the calibration of S11 at the front surface of a material for a coaxial-feed type cut-off circular waveguide with three reference materials inserted and no short termination condition was proposed as a preliminary step for dielectric measurement in liquids. The equations for jig calibration of S11 with these reference materials were first defined, and the electrostatic capacitance for the analytical model unique to the jig was quantified by substituting the reflection constant (calculated at frequencies of 0.50, 1.5 and 3.0 GHz using the mode-matching (MM) technique) into the equivalent circuit, assuming the sample liquid in the jig. The accuracy of S11 measured using the proposed method was then verified. S11 for the front surface of the sample material was also measured with various liquids in the jig after calibration, and the dielectric constants of the liquids were estimated as an inverse problem based on comparison of S11 calculated from an analytical model using EM analysis via the MM technique with the measured S11 values described above. The effectiveness of the proposed S11 calibration method was verified by comparison with dielectric constants estimated after S11 SOM (short, open and reference material) calibration and similar, with results showing favorable agreement with each method.
Kouji SHIBATA
Hachinohe Institute of Technology
The copyright of the original papers published on this site belongs to IEICE. Unauthorized use of the original or translated papers is prohibited. See IEICE Provisions on Copyright for details.
Copy
Kouji SHIBATA, "S11 Calibration of Cut-Off Circular Waveguide with Three Materials and Related Application to Dielectric Measurement for Liquids" in IEICE TRANSACTIONS on Electronics,
vol. E104-C, no. 2, pp. 93-101, February 2021, doi: 10.1587/transele.2020ECP5025.
Abstract: A method for the calibration of S11 at the front surface of a material for a coaxial-feed type cut-off circular waveguide with three reference materials inserted and no short termination condition was proposed as a preliminary step for dielectric measurement in liquids. The equations for jig calibration of S11 with these reference materials were first defined, and the electrostatic capacitance for the analytical model unique to the jig was quantified by substituting the reflection constant (calculated at frequencies of 0.50, 1.5 and 3.0 GHz using the mode-matching (MM) technique) into the equivalent circuit, assuming the sample liquid in the jig. The accuracy of S11 measured using the proposed method was then verified. S11 for the front surface of the sample material was also measured with various liquids in the jig after calibration, and the dielectric constants of the liquids were estimated as an inverse problem based on comparison of S11 calculated from an analytical model using EM analysis via the MM technique with the measured S11 values described above. The effectiveness of the proposed S11 calibration method was verified by comparison with dielectric constants estimated after S11 SOM (short, open and reference material) calibration and similar, with results showing favorable agreement with each method.
URL: https://global.ieice.org/en_transactions/electronics/10.1587/transele.2020ECP5025/_p
Copy
@ARTICLE{e104-c_2_93,
author={Kouji SHIBATA, },
journal={IEICE TRANSACTIONS on Electronics},
title={S11 Calibration of Cut-Off Circular Waveguide with Three Materials and Related Application to Dielectric Measurement for Liquids},
year={2021},
volume={E104-C},
number={2},
pages={93-101},
abstract={A method for the calibration of S11 at the front surface of a material for a coaxial-feed type cut-off circular waveguide with three reference materials inserted and no short termination condition was proposed as a preliminary step for dielectric measurement in liquids. The equations for jig calibration of S11 with these reference materials were first defined, and the electrostatic capacitance for the analytical model unique to the jig was quantified by substituting the reflection constant (calculated at frequencies of 0.50, 1.5 and 3.0 GHz using the mode-matching (MM) technique) into the equivalent circuit, assuming the sample liquid in the jig. The accuracy of S11 measured using the proposed method was then verified. S11 for the front surface of the sample material was also measured with various liquids in the jig after calibration, and the dielectric constants of the liquids were estimated as an inverse problem based on comparison of S11 calculated from an analytical model using EM analysis via the MM technique with the measured S11 values described above. The effectiveness of the proposed S11 calibration method was verified by comparison with dielectric constants estimated after S11 SOM (short, open and reference material) calibration and similar, with results showing favorable agreement with each method.},
keywords={},
doi={10.1587/transele.2020ECP5025},
ISSN={1745-1353},
month={February},}
Copy
TY - JOUR
TI - S11 Calibration of Cut-Off Circular Waveguide with Three Materials and Related Application to Dielectric Measurement for Liquids
T2 - IEICE TRANSACTIONS on Electronics
SP - 93
EP - 101
AU - Kouji SHIBATA
PY - 2021
DO - 10.1587/transele.2020ECP5025
JO - IEICE TRANSACTIONS on Electronics
SN - 1745-1353
VL - E104-C
IS - 2
JA - IEICE TRANSACTIONS on Electronics
Y1 - February 2021
AB - A method for the calibration of S11 at the front surface of a material for a coaxial-feed type cut-off circular waveguide with three reference materials inserted and no short termination condition was proposed as a preliminary step for dielectric measurement in liquids. The equations for jig calibration of S11 with these reference materials were first defined, and the electrostatic capacitance for the analytical model unique to the jig was quantified by substituting the reflection constant (calculated at frequencies of 0.50, 1.5 and 3.0 GHz using the mode-matching (MM) technique) into the equivalent circuit, assuming the sample liquid in the jig. The accuracy of S11 measured using the proposed method was then verified. S11 for the front surface of the sample material was also measured with various liquids in the jig after calibration, and the dielectric constants of the liquids were estimated as an inverse problem based on comparison of S11 calculated from an analytical model using EM analysis via the MM technique with the measured S11 values described above. The effectiveness of the proposed S11 calibration method was verified by comparison with dielectric constants estimated after S11 SOM (short, open and reference material) calibration and similar, with results showing favorable agreement with each method.
ER -