The search functionality is under construction.

IEICE TRANSACTIONS on Communications

Electric-Field Distribution Estimation in a Train Carriage due to Cellular Radios in order to Assess the Implantable Cardiac Pacemaker EMI in Semi-Echoic Environments

Takashi HIKAGE, Toshio NOJIMA, Soichi WATANABE, Takashi SHINOZUKA

  • Full Text Views

    0

  • Cite this

Summary :

The electromagnetic field (EMF) distributions created inside a train carriage by the cellular radios of the passengers are analyzed and the impact their electromagnetic interference (EMI) on the implantable cardiac pacemakers is evaluated based upon the analysis results. Both computer simulations and experiments using 800 MHz and 2 GHz transmitters in an actual train carriage confirm that excessively high EMF, high enough to affect the normal functions of the pacemaker, does not occur inside the carriage provided the safe distance of 22 cm specified for pacemaker users is kept. A simplified histogram estimation method for electric field strength is newly developed to deal with the complicated EMF distributions. It allows the EMI risk to pacemakers by cellular radio transmission to be quantitatively evaluated. Methodologies are described first. Typical results of FDTD analysis and actual measurement data are then shown. Finally, considerations and conclusions are made.

Publication
IEICE TRANSACTIONS on Communications Vol.E88-B No.8 pp.3281-3286
Publication Date
2005/08/01
Publicized
Online ISSN
DOI
10.1093/ietcom/e88-b.8.3281
Type of Manuscript
Special Section PAPER (Special Section of 2004 International Symposium on Electromagnetic Compatibility)
Category
Biological Effects

Authors

Keyword